生化怎么通样本针全自动生化分析仪贝克曼AU680仪器报警列表中文报警翻译

新闻资讯2026-04-21 18:35:22

1、报警表3038Power Failure Detect电源岀现断电3039Reage nt Refrigerator Temp. out of Range试剂冰箱温度超出范围3212R1 Reage nt Run OutR1试剂耗尽3213R1 Reage nt EmptyR1试剂空3214Probe Deterge nt Short探针清洗液少了3215Probe Deterge nt Empty探针清洗液空了3216Reage ntProbeW2Deterge ntShort试剂针W2清洗液少了3217R2 Reage nt ShortR2试剂少了3218R2 Reage nt Run O

2、utR2试剂耗尽3219R2 Reage nt EmptyR2试剂瓶空3222R2 Probe W2 Deterge nt ShortR2探针W2清洗液少了3223Reage ntRefrigeratorCover Open试剂冰箱盖打开了3225STAT Table Large Cover Ope n急诊仓大盖子打开了3226STAT Table Small Cover Ope n急诊仓小盖子打开了3230Reage nt Bottle Switch试剂瓶改变3232Reage ntShort for RB/ACAL/QC试剂少,不足以定标、质控3233STAT Sampling Inter

3、rupt急诊仓取样中断3234Reage nt Refrigerator On Use试剂冰箱正在使用3235RB/ACAL Sample Prior Diluent Short空白/定标预稀释液缺少3240Check the Tip of Sample Probe检查样本针尖3241Check the Tip of Reage nt Probe检查实际针尖3243Sample Syri nge Prime In complete样本注射器灌注未完成3250ISE Cover Ope n电解质盖打开3251ISE Buffer Solution Short电解质缓冲液少3252ISE MID

4、Solutio n Short电解质中值液少3253ISE REF Solutio n Short电解质内标液少3254ISE Deterge nt Short电解质清洗液少3255ISE Serum Standard Solution H Short电解质血清高值标准液少3256ISE Serum Standard Solution L Short电解质血清低值标准液少3259ISE Electrode Check Liquid Na Short钠电极检查液少了3260ISE Electrode Check Liquid K Short钾电极检查液少了3265ISE MID Solutio

5、n Stability Error电解质中值效期错误3301Sample ID Barcode Error样本条码错误3302Reage nt ID Barcode Error试剂条码错误3500DI Water Empty去离子水桶空了3501Deterge nt (C oncen trated) Empty浓缩清洗液桶空了3502Deterge nt Empty清洗液空了3503Sample Probe (FA) Error (Up-Dow n)样本针上下错误3504Reage nt Probe (FD) Error (Up-Dow n)试剂针上下错误3506Deterge nt Over

6、flow清洗液溢出3509ISE Data Low电解质数据低3510Sample Probe Deterge nt Empty样本探针清洗液空了( W1)3511Sample Prior Diluent Run Out样品预稀释液耗尽3512Sample Prior Dilue nt Empty样品预稀释瓶空了3513R1 Reage nt Run OutR1试剂耗尽(不足)3514R1 Reage nt EmptyR1试剂瓶空了3515Reage nt Probe Deterge nt Empty试剂探针清洗液空了3516R2 Reage nt Run OutR2试剂耗尽(不足)3517R

7、2 Reage nt EmptyR2试剂瓶空了3519Reage nt Refrigerator Cover Open试剂冰箱盖打开3521STAT Table Large Cover Ope n急诊仓大盖子打开3522STAT Table Small Cover Ope n急诊仓小盖子打开3524ISE Cover Ope n电解质盖子打开3526Rack Tran sfer Cover Open架子传输部位盖子打开3533Rack Storage Un it Full架子回收单元满了3534Rack Storage Un it Full架子回收单元满了3535STAT Sampling I

8、nterrupt急诊台吸样中断3536Rack Kind Error架子类型错误3537Rack ID Barcode Error (A)架子条码错误3538Rack Supply Unit Cover Ope n架子申请单元的盖子打开了3539Sample Probe (FA) Error (Liquid Sen sor)样品针液体传感器错误3540Reage nt Probe (FD01) Error (Liquid Sen sor)试剂针液体传感器错误3542Reage nt Refrigerator Cover Open试剂冰箱盖打开3543Reage nt Refrigerator

9、Cover Open试剂冰箱盖打开3544R2 Reage nt Refrigerator Cover Ope nR2试剂冰箱盖打开3546STAT Table Large Cover Ope n样本针向下吸样时急诊台大盖 子打开3547STAT Table Large Cover Ope n样本针旋转时急诊台大盖子打 开3548ISE Cover Ope n样本针旋转时ISE盖子打开3549ISE Cover Ope n分析仪初始化时ISE盖子被打开3550ISE A/D Unit ErrorISE进行反应时,A/D信号转化发 生错误3551DI Water Supply Un availa

10、ble(去离子水桶的感应阀出现错 误)水位达不到3601Rack Tran sfer Cover Open架子传送带的盖子打开3604Un removal Rack in Rack Supply Unit进架单元架子不能动3605Un removal Rack on Rack Tran sfer (A)架子传送单元架子不能动3606Un removal Rack in Rack Buffer Uni t (A)缓冲单元进架错误,即架子不能 动3607Un removal Rack in Repeat Run Li ne (A)复查单元,架子无法动3608Rack Storage Un it F

11、ull架子回收单元满了()3641Rack Supply Un it (CA) Error (A)A, B, C, D架子单元发生错误3642Rack Supply Belt (CB) Error (A)A, B, C, D架子传送带运行中发生错误3663RB In complete空白定标未完成3664ACAL In complete定标分析未完成3665QC In complete质控未完成5002INITIALIZE初始化5003WARM UP预热5004STANDBY待机5005MEASURE START测量开始5006NO MORE SAMPLE RACKS没有样本架5007MOVI

12、NG TO PAUSE转换成暂停5008MOVING TO STOP MODE转换成停止5009PAUSE暂停5010STOP停止5011SYSTEM END系统停止5014PHOTOCAL START光电检查开始5017PHOTOCAL END光电校正结束5018PHOTOCAL FAILURE光电校正失败5031INDEX UPDATED校正索引5032INDEX DELETED删除索引5033REACTION DATA DELETED删除反应数据5093PASSWORD CHANGE FAILURE (A)密码修改失败5101NO REAGENT BOTTLE没试剂瓶5103REAGEN

13、T ID READ ERROR试剂条码阅读错误5105REAGENT ID NOT DEFINED未定义试剂条码5107R2 REAGENT FOUND IN THE R1REFRIGERATOR在R1试剂仓发现R2试剂5108R1 REAGENT FOUND IN THE R2REFRIGERATOR (A, B, C)在R2试剂仓发现R1试剂5109ILLEGAL BOTTLE TYPE不规则的试剂瓶5110ILLEGAL REAGENT POSITION不规则的试剂位置5111REAGENT EXPIRED试剂失效5112DIFFERENT LOT NO. REAGENT FOUND51

14、14REAGENT TYPE CONVERTED试剂类型改变5115BOTTLE TYPE CONVERTED瓶子类型改变5201PHOTOCAL DATA ERROR光电数据错误5202CUVETTE MEAN CHECK ERROR比色杯均值检查错误5203PHOTOCAL LAMP CHECK ERROR光电灯泡检查错误5251INCUBATION BATH TEMPERATURE HIGH反应仓温度过高5252INCUBATION BATH TEMPERATURE LOW反应仓温度过低5253REAGENT REFRIGERATOR TEMPERATUREHIGH试剂冰箱温度过高530

15、1PANIC VALUE (A, B, C, D)达到病理值5302HbA1c PRETREATMENT CHECK RANGE ERROR糖化预处理范围检查错误5303HbA1c% CALCULATION FAILED糖化计算错误5304NO COLOR DATA无颜色实验数据5305NO BLANK DATA无空白实验数据5306RB DATA ERROR试剂空白数据错误5307CALIBRATION FACTOR RANGE OVER定标因数范围超上限5308CALIBRATION FACTOR RANGE UNDER定标因数范围超下限5309CALIBRATION OD RANGE O

16、VER定标吸光度范围超上限5310CALIBRATION OD RANGE UNDER定标吸光度范围超下限5311CALIBRATION ERROR定标错误5312MULTIPOINT-CAL. OD SLOPE CHECK ERROR多点定标吸光度斜率检查错误53131/2 POINT-CAL. CORRECTION FACTOR ERROR一点/两点校准因数错误5314QC DATA RANGE OVER质控数据范围超上限5315QC DATA RANGE UNDER质控数据范围超下限5317QC DATA DEVIATION CHECK ERROR质控数据产生偏差5401NO SAMPLE CUP (A, B-C, D)没有样本杯5402SAMPLE NOP (A, B-C, D)没有申请样本5403SAMPLE ID NOT FOUND未找到样本条码5404NO TESTS REQUISITIONED未申请测试5408SAMPLE ID READ ERROR样本条码阅读错误5751PRINTER NOT SELECTED未选择打印机5752PRINTER NOT AVAILABLE打印机不