| Output channels |
The 8 output channels are provided through a network as specified in IEC 60601-2-51 to provide signals to 10 lead electrodes; in the device under test, this will be displayed as 12 leads. |
8 outputs 10 lead electrodes 12 leads |
| Voltage accuracy |
IEC 60601-2-51 specifies a limit of ±1%, but does not provide a lower limit (all systems must have a lower limit). An inferred specification of 1% ±5?V is derived from the device under test specification in IEC 60601-2-51 of 5% ±25?V. |
±1% for values greater or equal to 500μV. ±5μV for values under 500μV.
|
| Output voltage |
The output voltage on most of the database
and ECG is +5mV~-5mV.
|
±5mV |
Output noise level 0-150Hz |
Output noise should not influence the test. A value a 5?V is suitable for this requirement. Can be verified by monitoring the signal in the device under test using a “diagnostic” filter setting. |
<5?V |
| Time accuracy |
IEC 60601-2-51 does not provide any limits. An inferred limit from the device under test. An inferred limit of ±1% is used (see 4.2). The system’s design accuracy exceeds 0.1% as a 100ppm crystal reference is used. |
±1% |
| Sampling rate |
A maximum sampling rate of 1kHz matches the sampling rates of ECG files. |
1kHz (8 channels) |
| Power supply |
Powered from the USB supply (5V 0.2A) |
N/A |
| Environment |
Intended for normal laboratory environment. The selection of critical components such as reference voltages, DAC, precision resistors are known to be stable in the range shown. |
15-30°C 10-95% RH
|